BS ISO 15708-3:2017 pdf download

admin
BS ISO 15708-3:2017 pdf download

BS ISO 15708-3:2017 pdf download.Non-destructive testing — Radiation methods for computed tomography — Part 3: Operation and interpretation
4.2 CT system set-up
4.2.1 General The CT system set-up is oriented towards the requirements for the given task. The required spatial resolution (taking into account the tube focal spot size), contrast resolution, voxel size and the CT image quality can be derived from these requirements. The quality of the CT image is determined by different parameters, which under certain circumstances counteract each other. In the following, system parameters are described and information is provided on setting up a CT system for inspection. Due to the interactions of the different system parameters, it may be necessary to run through the set-up steps several times in order to acquire optimal data. The optimal energy is that which gives the best signal-to-noise ratio and not necessarily that which gives the clearest radiograph (the dependency of the detector efficiency on the energy is to be taken into account). However, in order to differentiate between materials of different chemical composition it may be necessary to adjust the accelerating voltage to maximise the difference in their linear attenuation coefficients.
4.2.2 Geometry
The source-detector and source-object distances and thus also the beam angle used should be specified. In order to achieve high resolutions, the projection can be magnified onto the detector. The magnification is equal to the ratio of the source-detector distance to the source-object distance. Increasing source-detector distance leads to a reduced intensity at the detector and thus to a reduced signal to noise ratio. Accordingly, this also applies when using detectors with improved detector resolution, which can result in a reduction of the signal-to-noise ratio due to the reduced intensity per pixel. In general, for this reason, minimisation of the source-object distance is to be preferred.In order to obtain high beam intensity at the detector, the source-detector distance should be selected so that it is as small as possible taking into account the required resolution so that the beam cone still fully illuminates the detector. In the case of 3D-CT, the (in general vertical) total cone beam angle measured parallel to the rotation axis should typically be less than 15°, but this is specimen dependant, in order to minimise reconstruction-determined (Feldkamp) distortions of the 3D model. In addition, these restrictions do not apply for the perpendicular (in general horizontal) beam angle. For a higher geometric magnification, the object must be positioned as near as possible to the source, taking into consideration the limit on sharpness imposed by focal spot size. The rotation of the object must take place at at least 180 °plus beam angle of the X-ray beam, whereby an improved data quality is the result of an increasing number of angular increments. For this reason, the object is typically turned through 360 °. Ideally, the number of angular increments should be at least π/2 × matrix size (uneven number of projections per 360°) where the matrix size is the number of voxels across the sample diameter or the largest dimension. For more information, refer to 5.5.